Electrial Interface
Nurse call, USB, NET, RS232, VGA
Interface Type
Main patient outlet:ISO 22M/15F
Oxygen inlet:DISS or NIST male
Display
Display size: 15.6 in diagonal
Display type: Active matrix color TFT LCD
Display resolution: SVGA resolution, 1366 x 768
Controls
Fully electronic control
Ventilation type: Blower
Ventilation modes:
A/C-VCV, A/C-PCV, A/C-PRVC, SIMV-VC, SIMV-PC, SIMV-PRVC,
BIPAP, APRV, CPAP,CPAP-VG, NIV, IV.
Special functions:
manual breath, O2 therapy, standby, screen-lock, inspiratory hold,
expiratory hold, sigh,
ATC, PCVR, PPS, Compliance, p0.1, calibrate, nebulizer, O2 suction.
Patient types: Adult/Pediatric/Neonate
Patient height: Adjustable
Electrical and gas supplies
Input voltage:
100 to 240V ~ ±10 %, 50 / 60 Hz AC or 24 VDC±10 %
Power consumption: 50 W typical
Protection class: Class I
Backup battery:
2 batteries(5200mAh each)/ hot swappable
Oxygen supply:
280 to 600 kPa (41 to 87 psi), 120 L / min
Low pressure:
oxygen ≤15 L / min, max. 600 hPa for low pressure
Alarms
Operator-adjustable: Low/high minute volume, low/high pressure,
low/high tidal volume, low/high rate, apnea time, low/high oxygen
Optional: low/high CO2, low/high SPO2
Special alarms: Oxygen concentration, disconnection, loss of PEEP,
exhalation obstruction, flow sensor, power supply batteries, gas supply
Loudness Adjustable: 1 to 9
Event log: Storage and display of up to 1000 events with date and
time stamp
Physical dimensions
Weight: < 45.3 kg
Dimensions(WxHxD): Approx. 600x402x415mm